SPECIFIC MASS FLOW RATE MEASUREMENTS IN A PULSATING FLOW OF GAS

Journal Title: Metrology and Measurement Systems - Year 2008, Vol 15, Issue 2

Abstract

The method of measurement of the specific mass flow rate in a pulsating flow by means of the Constant Temperature Anemometer (CTA) has been described. Special attention has been paid to the probe calibration problem. Different forms of the CTA characteristic have been analyzed. Example measurement results have been presented for chosen pulse frequencies. Apart from the measurements executed in one single point representing each control section, a flow field survey has been done in order to determine the velocity profile under the conditions of a pulsating flow. Probes have been displaced radially with a small step to cover the range from the pipe axis to its wall. It has been found that in the large field around the pipe axis, successive velocity plots are similar as far as their shape and phase shift are considered. Pulsations are damped and mean velocity decreases rapidly only in direct proximity of the pipe wall. It has been also shown that the presence of pulsations makes the mean velocity profile more uniform than in the case of a steady flow. Measurements have been also performed for the case of a reverse flow occurring in the pipe at resonance frequencies.

Authors and Affiliations

ALEKSANDER OLCZYK

Keywords

Related Articles

EVALUATION OF CMM FOR FLATNESS MEASUREMENTS

Coordinate measuring machine (CMM) is a universal measuring instrument in dimensional metrology. However, the CMM has many sources of errors and deviations that affect its precision and accuracy. Unfortunately, the accur...

A METHOD OF IDENTIFICATION OF RTS COMPONENTS IN NOISE SIGNALS

Noise signals containing two components with different distribution of current values of the noise signal, i.e. a component with Gaussian distribution and a component with non-Gaussian distribution, were analyzed. The no...

THE SOURCE OF ERRORS IN THE OSCILLOMETRIC BLOOD PRESSURE MEASUREMENTS

This report presents an analysis of the source of errors in the blood pressure measurements which use the oscillometric method. The basis of this method are the blood pressure oscillations in the occluding cuff which are...

FREQUENCY-RESPONSE COMPENSATION OF DISTORTING PARTS OF MEASUREMENT SYSTEMS

In this article a class of digital correction filters is presented. This class of filters has been synthesized in an optimization procedure where two criteria are considered. The first criterion is an approximation of th...

PRE-PROCESSING AND PATTERN RECOGNITION METHODS FOR ARTIFICIAL OLFACTION SYSTEMS: A REVIEW

Data analysis is a fundamental part of artificial olfaction systems. Nonetheless, while electrical and mechanical components of electronic noses are deeply investigated and optimised for the specific applications, data a...

Download PDF file
  • EP ID EP144600
  • DOI -
  • Views 62
  • Downloads 0

How To Cite

ALEKSANDER OLCZYK (2008). SPECIFIC MASS FLOW RATE MEASUREMENTS IN A PULSATING FLOW OF GAS. Metrology and Measurement Systems, 15(2), 165-176. https://europub.co.uk./articles/-A-144600