A comparison of the shear bond strength of metal orthodontic brackets that have been treated with various light-emitting diode intensities and curing times: In-vitro research

Journal Title: International Journal of Experimental Research and Review - Year 2023, Vol 34, Issue 5

Abstract

The study aims to test the effects of LED light at three different power settings (500mW/cm2, 1000mW/cm2& 1500m W/cm2) and three different exposure times (5sec, 10sec & 20sec) adhesive composites for use with stainless steel orthodontic brackets, and their shear bond strengths. Current In-vitro work used 120 teeth, mostly premolars, separated into 10 equal groups of 12. After the crowns of the teeth were exposed by mounting them vertically into self-heal acrylics, the buccal-enamel exterior was prepared for bonding (PEA) 0.022”x 0.028” stainless steel premolar bracket using a thin layer of light cure adhesive primer and paste (Transbond-XT-3M Unitek Monrovia, Calif). Brackets in Group 1 comprising the Control group treated by usual halogen light source for forty sec, in remaining 9 groups (Experimental group) three commercially available LED curing light units i.e., Woodpecker, Guilin, Guangxi, China; Stealth Soft Equinox, Denmark, Holland; Radii plus, SDI, Australia, having three different light intensity 500mW/cm2, 1000mW/cm2, 1500mW/cm2 respectively were used to bond the brackets at three contact time (5/10/20 sec). The universal testing equipment was used to determine the shear bond strength of each specimen. The uppermost average SBS was found in the group cured with LED of intensity 1500 mW/cm2 for 20 sec (20.08 + 4.82 MPa) which was statistically significant compared the to control group (12.72 + 2.42 MPa). The lowest mean SBS values were obtained for groups cured with LED of intensity 500 mW/cm2 even at different curing times of 5, 10 and 20 sec i.e., 5.46 + 1.99 MPa, 8.77 + 2.38 Mpa and 11.24 + 2.59 MPa, respectively. Increasing the intensity of the LED light curing unit provides a clinically acceptable SBS even after reducing the curing times, thereby reducing the chairside time for bonding of orthodontic brackets.The study aims to test the effects of LED light at three different power settings (500mW/cm2, 1000mW/cm2& 1500m W/cm2) and three different exposure times (5sec, 10sec & 20sec) adhesive composites for use with stainless steel orthodontic brackets, and their shear bond strengths. Current In-vitro work used 120 teeth, mostly premolars, separated into 10 equal groups of 12. After the crowns of the teeth were exposed by mounting them vertically into self-heal acrylics, the buccal-enamel exterior was prepared for bonding (PEA) 0.022”x 0.028” stainless steel premolar bracket using a thin layer of light cure adhesive primer and paste (Transbond-XT-3M Unitek Monrovia, Calif). Brackets in Group 1 comprising the Control group treated by usual halogen light source for forty sec, in remaining 9 groups (Experimental group) three commercially available LED curing light units i.e., Woodpecker, Guilin, Guangxi, China; Stealth Soft Equinox, Denmark, Holland; Radii plus, SDI, Australia, having three different light intensity 500mW/cm2, 1000mW/cm2, 1500mW/cm2 respectively were used to bond the brackets at three contact time (5/10/20 sec). The universal testing equipment was used to determine the shear bond strength of each specimen. The uppermost average SBS was found in the group cured with LED of intensity 1500 mW/cm2 for 20 sec (20.08 + 4.82 MPa) which was statistically significant compared the to control group (12.72 + 2.42 MPa). The lowest mean SBS values were obtained for groups cured with LED of intensity 500 mW/cm2 even at different curing times of 5, 10 and 20 sec i.e., 5.46 + 1.99 MPa, 8.77 + 2.38 Mpa and 11.24 + 2.59 MPa, respectively. Increasing the intensity of the LED light curing unit provides a clinically acceptable SBS even after reducing the curing times, thereby reducing the chairside time for bonding of orthodontic brackets.

Authors and Affiliations

Gananjay Patil, Santosh Kumar, Namrata Dogra, Jasmine Nindra, Satwashila Kadam, Krishna Yadav

Keywords

Related Articles

Exploration of ethno-medicinal herbs and their practices by indigenous people of Achanakmar regions of Chhattisgarh State, India

The study was performed in the Achanakmar regions of Chhattisgarh state, India. At the study site, a total of 54 herbaceous medicinal plants belonging to 30 families were documented. Between March 2020 and March 2022, i...

A Noise reduction in the medical images using hybrid combination of filters with nature-inspired Black Widow Optimization Algorithm

This paper proposes an image filtering method to remove the noises in medical images in a controlled manner. To achieve this goal, the optimal parameters of the conventional filters are determined using the nature-inspir...

Kruskal Wallis and mRMR Feature Selection based Online Signature Verification System using Multiple SVM and KNN

Signature verification is a very important research area. Signature has been widely accepted as a person authentication method for centuries. It is mostly used in financial transactions, document authentication and agree...

Benefits of Yogic Practice on Body Fat Composition in Obese Adults

Obesity is a worldwide concern and is increasing rapidly across the continents. Yoga can be used as a preventive method for obesity. Therefore, current research investigates the effects of a yoga intervention on body com...

Development and Validation of RP-HPLC Method for Estimation of Ticagrelor in Pharmaceutical Dosage Form and Force degradation study

Ticagrelor is a selective Adenosine diphosphate (ADP)-receptor antagonist which is prescribed in the form of tablets and acts as an oral antiplatelet for the prevention of further thrombotic events in patients with Acute...

Download PDF file
  • EP ID EP722896
  • DOI https://doi.org/10.52756/ijerr.2023.v34spl.001
  • Views 39
  • Downloads 0

How To Cite

Gananjay Patil, Santosh Kumar, Namrata Dogra, Jasmine Nindra, Satwashila Kadam, Krishna Yadav (2023). A comparison of the shear bond strength of metal orthodontic brackets that have been treated with various light-emitting diode intensities and curing times: In-vitro research. International Journal of Experimental Research and Review, 34(5), -. https://europub.co.uk./articles/-A-722896