ASSESSMENT OF THE FACTORS INFLUENCING AND COMPARING THE INTRAOCULAR PRESSURE WITH THE HELP OF SCHIOTZ INDENTATION TONOMETER AND GOLDMANN’S APPLANATION TONOMETER: A CLINICAL STUDY
Journal Title: International Journal of Medical Research & Health Sciences (IJMRHS) - Year 2015, Vol 4, Issue 1
Abstract
Background: Intraocular pressure (IOP) is the fluid pressure inside the eye and is maintained by the equilibrium between the forces determining the formation of aqueous humour and the alteration in the resistance to its outflow. IOP is important to evaluate patients who are at an increased risk of glaucoma. Clinically measurement of IOP is based on principles of indentation and applanation and such a method is called Tonometry. The tonometers used today are the Schiotz Indentataion Tonometer (ST) and the Goldmann’s Applanation Tonometer (GAT). However the latter is accepted as the standard one. Aim: 1) To estimate the normal mean IOP for the population under study.2) To study the various factors- age, sex, refractive errors that influence the IOP.3) Merits and demerits of the individual tonometers. Method: A sample size of 100 cases of 200 sets of eyes was taken and divided into 2 groups, Group A with emmetropia, myopia, hypermetropia of both sexes and Group B with frank glaucomatous changes of both sexes. 3 consecutive measurements with GAT were recorded in each eye followed by ST with 5.5gm weight first followed by ST with 7.5gm weight. The whole database was recorded and statistically analysed. Results: Out of the total 200 eyes studied, Group A consisted of 168 apparently normal eyes which included emmetropes, myopes and hypermetropes and Group B comprised of 32 eyes which were frank glaucomatous cases with glaucomatous field defects. On estimating the mean IOP with the help of GAT and ST between the 2 eyes there was not a significant difference. On comparing the refractive status, myopes showed a higher IOP than hypermetropes and emmetropes. With both GAT and ST females had higher IOP than males. The mean IOP increased as age progressed. Group B studies which had frank glaucoma cases showed that the IOP measured with GAT was higher and more accurate to the actual IOP value whereas the readings with Schiotz were variable and unreliable. Conclusion: IOP is one of the key metrics which is used to monitor the health of one’s eye especially an eye with glaucoma. The IOP measured with GAT was more accurate towards the true intraocular pressure and hence Applanation Tonometer is considered gold standard in measuring IOP.
Authors and Affiliations
Mridula V Amarnath| Postgraduate, Department of Ophthalmology, Meenakshi Medical College, Kanchipuram, Tamil Nadu, India, Corresponding author email: mridulavenugopal.88@gmail.com, Christina Samuel| Postgraduate, Department of Ophthalmology, Meenakshi Medical College, Kanchipuram, Tamil Nadu, India, Sundararajan D| Professor & Head, Department of Ophthalmology, Meenakshi Medical College, Kanchipuram, Tamil Nadu, India
The effect of electronic education on interest of learning Students of Payame Noor University
This research aimed to investigate the impact of electronic education of Payame Noor University Students interest to learning. For this purpose, were used the experimental method and pretest and post-test. For this purp...
Rapid magnetic solid phase extraction–spectrophotometric determination of doxycycline and erythromycin in real samples with nanoparticles
In recent years, pharmaceutical and personal care products (PPCPs) have been detected in diverse environments (including groundwater, river water, and municipal wastewater).This work characterizes removal of trace anti...
STUDENT EVALUATION OF AN OSCE IN GENERAL MEDICINE AT MAMATA MEDICAL COLLEGE, ANDHRA PRADESH
The assessment of student’s clinical competence is of paramount importance, and there are several means of evaluating student performance in medical examinations. The OSCE is an approach to student assessment in...
Relationship between type of nutrition and infantile colic
Background: Infantile colic is one of the most common causes of mothers’ referral to physicians and pediatricians for treatment of their infants. This disorder is characterized as unnatural excessive crying which has...
The comparison of the impact of lavender and Valerian aromatherapy on reduction of the active phase among Nulliparous women: A double blind randomized controlled trial
One of the important points in health care of nulliparous women is reduced pain, anxiety coping and reduced delivery time. Aromatherapy is one of the effective non-pharmacological methods. The present study is aimed to...