Deposition Pattern of Polydisperse Dry Powders in Andersen Cascade Impactor - Aerodynamic Assessment for Inhalation Experimentally and In Silico

Journal Title: Turkish Journal of Pharmaceutical Sciences - Year 2020, Vol 17, Issue 1

Abstract

Objectives: An impactor is a standard instrument that applied for particle deposition assessment in the pharmaceutical aerosols. It provides data comparison between inhaler formulations. However, the deposition pattern in the impactor is not clearly understood. In practice monodisperse aerosols were employed to calibrate the impactor. Materials and Methods: This study used polydisperse aerosols together with the computer simulation to track the particles in the impactor to understand the deposition pattern. Particles deposited on each stage of the Andersen cascade impactor were compared with its stage cut-off diameter using polydisperse aerosols by three particle sizing techniques. The relationship of cut-off diameter with particle size distribution was established for each stage. Also, the computational verification was used to complement the real experiments. Results: Projected diameters from microscope images showed that the size of particles varied on the stage’s collection plate, and the median size of each stage decreased along the lower stages from 8.53 to 0.92 μm. The median sizes measured by laser diffraction were close to the impactor’s cut-off diameters. In silico data showed that the outlet mass fractions gradually changed in size towards the lower stages. Conclusion: Polydisperse aerosols and in silico computer fluid dynamics may compliment to standard calibration method.

Authors and Affiliations

Janwit DECHRAKSA, Tan SUWANDECHA, Teerapol SRICHANA

Keywords

Related Articles

Development and Optimization of a Floating Multiparticulate Drug Delivery System for Norfloxacin

Objectives: Norfloxacin is a synthetic broad-spectrum antibacterial drug having poor bioavailability and pH-dependent solubility. The purpose of the present study was to develop a gastroretentive floating multiparticulat...

In Vitro Skin Permeation and Antifungal Activity of Naftifine Microemulsions

Objectives: Microemulsions are fluid, isotropic, colloidal systems that have been widely studied as drug delivery systems. The percutaneous transport of active agents can be enhanced by their microemulsion formulation wh...

Investigation of Physiological Effects Induced by Dehydroepiandrosterone in Human Endothelial Cells and Ovarian Cancer Cell Line

Objectives: Dehydroepiandrosterone (DHEA) is an endogenous hormone that acts as a ligand for several cellular receptors. An age-dependent decline in circulating levels of DHEA is linked to changes in various physiologica...

Investigation of the Voltammetric Behavior of Methyldopa at a Poly (p-Aminobenzene Sulfonic Acid) Modified Sensor

Objectives: The aim was to modify carbon electrodes with (p-aminobenzene sulfonic acid) and use them as a sensor for sensitive and reliable detection of methyldopa (MD) and ascorbic acid. Materials and Methods: Electrop...

Evaluation of the Methylation and Acetylation Profiles of Dinitroaniline Herbicides and Resveratrol on the V79 Cell Line

Objectives: Herbicides are among the most widely used pesticide compounds for plant growth control worldwide. Risk assessment of the dinitroaniline-derived herbicides pendimethalin and trifluralin is important for foodbo...

Download PDF file
  • EP ID EP684725
  • DOI 10.4274/tjps.galenos.2018.15870
  • Views 172
  • Downloads 0

How To Cite

Janwit DECHRAKSA, Tan SUWANDECHA, Teerapol SRICHANA (2020). Deposition Pattern of Polydisperse Dry Powders in Andersen Cascade Impactor - Aerodynamic Assessment for Inhalation Experimentally and In Silico. Turkish Journal of Pharmaceutical Sciences, 17(1), -. https://europub.co.uk./articles/-A-684725