Doxorubicin hydrochloride. Determination in workplace air with high performance liquid chromatography
Journal Title: Podstawy i Metody Oceny Środowiska Pracy - Year 2018, Vol 34, Issue 2
Abstract
This article presents a method for measuring doxorubicin hydrochloride in workplace air with HPLC with diode array detector (DAD). The method is based on adsorption inhalable fraction of doxorubicin hydrochloride aerosol on glass fiber filter, desperation with water, and chromatographic analysis. The analysis was performed in reverse phase (mobile phase – 0.05 mol/L hydrophosphate disodum and acetonitrile (65:35) with pH – 3 with 0.5 mL/L triethylamine) on C18 column. The measurement range was 0.06 – 1 μg/m3 for 4800-L air sample. The limit of detection (LOD) was 0,0005 μg/ml and the limit of quantification (LOQ) was 0,0015 μg /ml. The developed method of doxorubicin hydrochloride determination has been recorded as an analytical procedure (see appendix).<br/><br/>
Authors and Affiliations
Małgorzata Szewczyńska
tert-Butyl ethyl eter
Znowelizowana metoda oznaczania chlorowodoru w powietrzu na stanowiskach pracy
Metodę stosuje się do oznaczania chlorowodoru w powietrzu na stanowiskach pracy.Metoda polega na przepuszczeniu znanej objętości badanego powietrza przez filtr bibułowy impregnowany węglanem sodu. Powstałą na filtrze sól...
Trimetyloamina. Dokumentacja proponowanych dopuszczalnych wielkości narażenia zawodowego
Trimetyloamina (TMA) w temperaturze pokojowej jest gazem palnym o bardzo nieprzyjemnym zapachu zepsutych ryb. Próg zapachowy trimetyloaminy znajduje się w przedziale 0,5 ÷ 1,9 μg/m3. Substancja ta bardzo dobrze rozpuszcz...
Ethyl acetate. Documentation
Ethyl acetate is colourless, transparent liquid with a fruity odour. It is one of the most widely used solvents and extractants in various industrial applications. Ethyl acetate is absorbed well by inhalation and rapidly...
Nitropropane. Determination in the workplace air with gas chromatography
A new procedure has been developed for the assay of 1-nitropropane and 2-nitropropane with gas chromatography with a flame ionization detector. This method is based on the adsorption of nitropropane vapors on Carbosieve...