Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography

Journal Title: Molecular Imaging and Radionuclide Therapy - Year 2019, Vol 28, Issue 3

Abstract

Objectives: The frequency of abdominal computed tomography examinations is increasing, leading to a significant level of patient dose. This study aims to quantify and evaluate the effects of automatic tube current modulation (ATCM) technique on patient dose and image quality in contrast-enhanced biphasic abdominal examinations. Methods: Two different scan protocols, based on constant tube current and ATCM technique, were used on 64 patients who visited our radiology department periodically. For three patient groups with different patient size, results from two protocols were compared with respect to patient dose and image quality. Dosimetric evaluations were based on the Computed Tomography Dose Index, dose length product, and effective dose. For the comparison of image qualities between two protocols, Noise Index (NI) and Contrast to Noise Ratio (CNR) values were determined for each image. Additionally, the quality of each image was evaluated subjectively by an experienced radiologist, and the results were compared between the two protocols. Results: Dose reductions of 31% and 21% were achieved by the ATCM protocol in the arterial and portal phases, respectively. On the other hand, NI exhibited an increase between 9% and 46% for liver, fat and aorta. CNR values were observed to decrease between 5% and 19%. All images were evaluated by a radiologist, and no obstacle limiting a reliable diagnostic evaluation was found in any image obtained by either technique. Conclusion: These results showed that the ATCM technique reduces patient dose significantly while maintaining a certain level of image quality.

Authors and Affiliations

Ayşegül Yurt, İsmail Özsoykal, Funda Obuz

Keywords

Related Articles

PET/CT Findings of a Patient with Cardiac Metastasis of Subungual Malign Melanoma

A 58-year old patient with a history of subungual malign melanoma was referred to our department for a 18F-FDG positron emission tomography (PET)/computed tomography (CT) whole body scan. An unexpected 18F-FDG uptake in...

18Fluorine-fluorodeoxyglucose PET/CT Imaging in Childhood Malignancies

Objectives: The aim of the study was to evaluate the utility of 18fluorine-fluorodeoxyglucose (18F-FDG) positron emission tomography/computed tomography (PET/CT) in the diagnosis, staging, restaging, and treatment respon...

Incidental “Urinoma” in 18F-FDG PET/CT Scan

18F-FDG PET/CT scanning was performed for the primary staging of a 47-year-old man with urothelial carcinoma. The patient underwent biopsy by ureteroscopy 15 days ago and the PET images revealed 18F-FDG accumulation in t...

Giant Abdominal Aortic Aneurysm in Bone Scan

Abdominal aortic aneurysm (AAA) may be incidentally detected in three-phased bone scintigraphy. AAA should be diagnosed prior to the development of symptoms to perform elective repair surgery. We present a rare case who...

68Ga PSMA Uptake at Roux-en-Y Eso-jejunostomy Junction Mimicking the Recurrence of Gastric Carcinoma in PET/CT

A 67-year-old male patient had undergone total gastrectomy and Roux-en-Y eso-jejunostomy 3 years ago for the treatment of tubular adenocarcinoma located at the corpus of the stomach. The patient was diagnosed with Gleaso...

Download PDF file
  • EP ID EP50510
  • DOI 10.4274/mirt.galenos.2019.83723
  • Views 273
  • Downloads 0

How To Cite

Ayşegül Yurt, İsmail Özsoykal, Funda Obuz (2019). Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography. Molecular Imaging and Radionuclide Therapy, 28(3), -. https://europub.co.uk./articles/-A-50510