Factors Affecting the Use of Impedance Spectroscopy in the Characterisation of the Freezing Stage of the Lyophilisation Process: the Impact of Liquid Fill Height in Relation to Electrode Geometry

Journal Title: AAPS PharmSciTech - Year 2014, Vol 15, Issue 2

Abstract

This study aims to investigate the application of impedance spectroscopy using fixed electrode geometries on a standard glass vial in the characterisation of the freezing process of solutions at different fill liquid volumes. Impedance spectra (between 10 and 106 Hz) were recorded every 3 min, during the freezing cycle on a solution of 30 mg/mL sucrose contained within 10 mL glass vials having an electrode system (two thin copper foils: w, 18 mm; h, 5 mm) affixed to the external surface of the vial. A fill factor (Φ) was defined in terms of the relative height of the solution volume to the height of the electrodes from the base of the vial. Solution volumes of 1.5 to 5 mL (corresponding to Φ= 0.5–1.6) were investigated to establish the applicability of having a fixed electrode geometry for a range of solution volumes. A linear relationship between the time duration of the ice formation/solidification phase and the fill factor suggests that fixed electrode geometries may be used to investigate a range of fill volumes. The benefit of this approach is that it does not invade the solution and hence records the freezing process without providing additional nucleation sites and in a manner which is representative of the entire fill volume.

Authors and Affiliations

Geoff Smith, Muhammad Sohail Arshad, Eugene Polygalov, Irina Ermolina

Keywords

Related Articles

Preparation of Effective and Safe Gene Carriers by Grafting Alkyl Chains to Generation 5 Polypropyleneimine

Gene therapy is a novel method to treat a variety of diseases including genetic disorders and cancer. Nonviral gene carriers have now gained considerable attention as gene carrier systems. Polyamidoamine (PAMAM) and poly...

Coatings of Eudragit® RL and L-55 Blends: Investigations on the Drug Release Mechanism

In a previous study, generally lower drug release rates from RL:L55 blend coated pellets in neutral/basic release media than in acidic release media were reported. The aim of this study was to obtain information on the d...

Strategies of Bringing Drug Product Marketing Applications to Meet Current Regulatory Standards

In the past decade, many guidance documents have been issued through collaboration of global organizations and regulatory authorities. Most of these are applicable to new products, but there is a risk that currently mark...

Application of Ethylcellulose Coating to Hydrophilic Matrices: A Strategy to Modulate Drug Release Profile and Reduce Drug Release Variability

The online version of this article (doi:10.1208/s12249-014-0128-5) contains supplementary material, which is available to authorized users.

The Anti-Melanoma Efficiency of the Intratumoral Injection of Cucurbitacin-Loaded Sustained Release Carriers: In Situ-Forming Implants

The online version of this article (doi:10.1208/s12249-015-0292-2) contains supplementary material, which is available to authorized users.

Download PDF file
  • EP ID EP682347
  • DOI  10.1208/s12249-013-0054-y
  • Views 96
  • Downloads 0

How To Cite

Geoff Smith, Muhammad Sohail Arshad, Eugene Polygalov, Irina Ermolina (2014). Factors Affecting the Use of Impedance Spectroscopy in the Characterisation of the Freezing Stage of the Lyophilisation Process: the Impact of Liquid Fill Height in Relation to Electrode Geometry. AAPS PharmSciTech, 15(2), -. https://europub.co.uk./articles/-A-682347