Neutron dose evaluation of Elekta Linac at two energies (10 & 18 MV) by MCNP code and comparison with experimental measurements
Journal Title: JOURNAL OF ADVANCES IN PHYSICS - Year 0, Vol 6, Issue 1
Abstract
Medical linear accelerators, besides the clinically high energy electron and photon beams, produce other secondary particles such as neutrons which escalate the delivered dose. In this study the neutron dose at 10 and 18MV Elekta linac was obtained by using TLD600 and TLD700 as well as Monte Carlo simulation. For neutron dose assessment in 2020 cm2 field, TLDs were calibrated at first. Gamma calibration was performed with 10 and 18 MV linac and neutron calibration was done with 241Am-Be neutron source. For simulation, MCNPX code was used then calculated neutron dose equivalent was compared with measurement data. Neutron dose equivalent at 18 MV was measured by using TLDs on the phantom surface and depths of 1, 2, 3.3, 4, 5 and 6 cm. Neutron dose at depths of less than 3.3cm was zero and maximized at the depth of 4 cm (44.39 mSvGy-1), whereas calculation resulted in the maximum of 2.32 mSvGy-1 at the same depth. Neutron dose at 10 MV was measured by using TLDs on the phantom surface and depths of 1, 2, 2.5, 3.3, 4 and 5 cm. No photoneutron dose was observed at depths of less than 3.3cm and the maximum was at 4cm equal to 5.44mSvGy-1, however, the calculated data showed the maximum of 0.077mSvGy-1 at the same depth. The comparison between measured photo neutron dose and calculated data along the beam axis in different depths, shows that the measurement data were much more than the calculated data, so it seems that TLD600 and TLD700 pairs are not suitable dosimeters for neutron dosimetry in linac central axis due to high photon flux, whereas MCNPX Monte Carlo techniques still remain a valuable tool for photonuclear dose studies.
Authors and Affiliations
Negin Shagholi, Hassan Ali, Mahdi Sadeghi, Arjang Shahvar, Hoda Darestani, Banaee Nooshin, Kheirolah Mohammadi
The Conservation of Energy and Momentum in the Positron-Electron Annihilation According to Complete Relativity
According to Complete Relativity (an improvement of Special Relativity where a new postulate was introduced: an electron cannot move at average speeds less than u0 = αc / Ï where α is the constant of fine structu...
Effect of UV Irradiation on The Structure and Optical Properties of PVA/Cucl2
The morphology and optical properties of doped PVA with Copper Chloride, CuCl2 have been investigated. The morphology of doped samples has been examined using both atomic force microscope (AFM) and scanning e...
Structure analysis and enhancement of creep resistance and thermal properties of eutectic Sn-Ag lead free solder alloy by Ti and Cd - addition
Eutectic (Sn-3.5wt.%Ag) solder alloy is used in electronic circuits in which the creep property of the solder joints is essential for their applications. The study of creep, structure and thermal properties of three sold...
Transition in spacetime
The universe has two main dimensions spatial dimension (consists of three dimensions directional X, Y, Z) and the other dimension is the temporal dimension. Time and space are linked strongly inseparable so we will consi...
Avoiding the Brain Damage
The pursuit of low losses and low distortion are necessary conditions to improve the signal integrity. As mentioned in [1], it is sometimes said that thereare two types of engineers” those who have signal integrityproble...