PROFICIENCY TESTING (PT) PROGRAM UNDER NABL IN THE PRESSURE RANGE 7- 70 MPA USING DIGITAL PRESSURE CALIBRATOR (DPC)

Journal Title: Metrology and Measurement Systems - Year 2005, Vol 12, Issue 3

Abstract

This paper describes the proficiency testing of the seven laboratories, accredited by National Accreditation Board for Testing and Calibration of Laboratories (NABL), India using a Digital Pressure Calibrator (DPC) [Model No. - H540/101, Sl. No.- MC808, make-DH-Budenberg, U.K.] as an artifact in the pressure range 70 bar to 700 bar (7 MPa to 70 MPa). The primary objective for organizing this proficiency testing (PT) is to assess the laboratory's technical competence to perform measurements and also fulfilling the requirement of ILAC/APLAC in regards to the compatibility of results submitted by these laboratories. National Physical Laboratory (NPLI), New Delhi, India has the responsibility of coordinating this programme and also acted as a reference laboratory. The stability of the artifact was assessed by measurements made at NPLI before and after the circulation of the artifact. The comparison was carried out at 11 arbitrarily chosen pressure points i.e. 0.7 MPa, 14 MPa, 21 MPa, 28 MPa, 35 MPa, 42 MPa, 49 MPa, 56 MPa, 63 MPa and 70 MPa throughout the entire pressure range. Six measurements were performed at each eleven pressure points, three each in increasing and decreasing orders of pressures, respectively. Laboratories were advised to report the values of various parameters related to the artifact including the temperatures at which the measurements were made. As it was a digital gauge, all participating laboratories were requested to perform the measurement at a temperature sufficiently close to 23°C so that the standard uncertainty of the various coefficient of the artifact do not contribute to the overall measurement uncertainty. Data are analyzed as per ISO/IEC GUM document. The normalized error (En) values are estimated in the entire pressure range. These results are quite encouraging for the manufacturers and calibration laboratories that are facing a new environment after the WTO agreement.

Authors and Affiliations

SANJAY YADAV, A. BANDYOPADHYAY

Keywords

Related Articles

LINEAR OPERATION OF DISCRETE SIGNAL TRACKING

The idea of discrete signal "tracking" action was put into practice many years ago. So-called dot recorders are best examples of its application. The dynamics of the dot-recorder measuring mechanism can be linear and non...

OPTICAL VORTICES AS PHASE MARKERS TO WAVE-FRONT DEFORMATION MEASUREMENT

n this paper a new method to the test deformation of the wave-front of the plane wave is presented. In the method Optical Vortex Interferometry is applied. In the OVI a regular lattice of optical vortices is generated by...

CONDUCTANCE QUANTIZATION IN FERROMAGNETIC CO NANOWIRES

Room temperature electronic transport properties of ferromagnetic quantum wires have not been yet fully understood, and the role of electronic structure of magnetic atoms in the conductance quantization is still under di...

PARTICULARITIES OF THE CYCLIC A/D CONVERTERS ENOB DEFINITION AND MEASUREMENT

The paper is devoted to the methods of accurate assessment and analysis of the expected and actual effective number of bits (ENOB) of cyclic analog-to-digital converters. The equivalence of definition of ENOB in the IEEE...

AN AUTOMATIC SYSTEM FOR IDENTIFICATION OF RANDOM TELEGRAPH SIGNAL (RTS) NOISE IN NOISE SIGNALS

In the paper the automatic and universal system for identi.cation of Random Telegraph Signal (RTS) noise as a non-Gaussian component of the inherent noise signal of semiconductor devices is presented. The system for data...

Download PDF file
  • EP ID EP123843
  • DOI -
  • Views 80
  • Downloads 0

How To Cite

SANJAY YADAV, A. BANDYOPADHYAY (2005). PROFICIENCY TESTING (PT) PROGRAM UNDER NABL IN THE PRESSURE RANGE 7- 70 MPA USING DIGITAL PRESSURE CALIBRATOR (DPC) . Metrology and Measurement Systems, 12(3), 323-337. https://europub.co.uk./articles/-A-123843