Quantitation of motexafin lutetium in human plasma by liquid chromatography-tandem mass spectrometry and inductively coupled plasma-atomic emission spectroscopy

Journal Title: The AAPS Journal - Year 2003, Vol 5, Issue 3

Abstract

Liquid chromatography-tandem mass spectrometry (LC-MS/MS) and inductively coupled plasma-atomic emission spectroscopy (ICP-AES) methods were developed and validated for the evaluation of motexafin lutetium (MLu, lutetium texaphyrin, PCI-0123) pharmacokinetics in human plasma. The LC-MS/MS method was specific for MLu, whereas the ICP-AES method measured total elemental lutetium. Both methods were fast, simple, precise, and accurate. For the LC-MS/MS method, a closely related analogue (PCI-0353) was used as the internal standard (IS). MLu and the IS were extracted from plasma by protein precipitation and injected onto and LC-MS/MS system configured with a C18 column and an electrospray interface. The lower limit of quantitation was 0.05 μg MLu mL−1, with a signal-to-noise ratio of 15∶1. The response was linear from 0.05 to 5.0 μg MLu mL−1. For the ICP-AES method, indium was used as the IS. The sample was digested with nitric acid, diluted, filtered, and then injected onto the ICP-AES system. Two standard curve ranges were validated to meet the expected range of sample concentrations: 0.5 to 50, and 0.1 to 10 μg Lu mL−1. The LC-MS/MS and ICP-AES methods were validated to establish accuracy, precision, analyte stability, and assay robustness. Interday precision and accuracy of quality control samples were ≤6.3% coefficient of variation (CV) and within 2.2% relative error (RE) for the LC-MS/MS method, and ≤8.7% CV and within 4.9% RE for the ICP-AES method. Plasma samples from a subset of patients in a clinical study were analyzed using both methods. For a representative patient, over 90% of the elemental lutetium in plasma could be ascribed to intact MLu at early time points. This percentage decreased to 59% at 48 hours after dosing, suggesting that some degradation and/or metabolism of the drug may have occurred.

Authors and Affiliations

Dale Miles, Tarak D. Mody, Lori I. Hatcher, John Fiene, Mark Stiles, Patrick P. Lin, J. W. Lee

Keywords

Related Articles

Best practices during bioanalytical method validation for the characterization of assay reagents and the evaluation of analyte stability in assay standards, quality controls, and study samples

Characterization of the stability of analytes in biological samples collected during clinical studies together with that of critical assay reagents, including analyte stock solutions, is recognized as an important compon...

Impact of Anti-Drug Antibodies in Preclinical Pharmacokinetic Assessment

The administration of human biotherapeutics is often associated with a higher incidence of immunogenicity in preclinical species. The presence of anti-drug antibodies (ADAs) in the test samples can affect the accurate me...

Drug discovery from natural sources

Organic compounds from terrestrial and marine organisms have extensive past and present use in the treatment of many diseases and serve as compounds of interest both in their natural form and as templates for synthetic m...

Simplification of Complex Physiologically Based Pharmacokinetic Models of Monoclonal Antibodies

The online version of this article (doi:10.1208/s12248-014-9591-x) contains supplementary material, which is available to authorized users.

Novel system to investigate the effects of inhaled volume and rates of rise in simulated inspiratory air flow on fine particle output from a dry powder inhaler

This study evaluated the effect of inhaled volume and simulated inspiratory flow rate ramps on fine particle output from dry powder inhalers (DPIs). A simple, robust system was developed to account for “rate of r...

Download PDF file
  • EP ID EP681969
  • DOI  10.1208/ps050323
  • Views 100
  • Downloads 0

How To Cite

Dale Miles, Tarak D. Mody, Lori I. Hatcher, John Fiene, Mark Stiles, Patrick P. Lin, J. W. Lee (2003). Quantitation of motexafin lutetium in human plasma by liquid chromatography-tandem mass spectrometry and inductively coupled plasma-atomic emission spectroscopy. The AAPS Journal, 5(3), -. https://europub.co.uk./articles/-A-681969