STUDY OF SOME ESSENTIAL QUALITY CONTROL TEST PARAMETERS FOR THE PERFORMANCE OF DOSE CALIBRATORS IN NUCLEAR MEDICINE PRACTICES.

Journal Title: International Journal of Advanced Research (IJAR) - Year 2019, Vol 7, Issue 3

Abstract

The aim of this study is to evaluate and compare the performance of two dose calibrators which are used in Nuclear Medicine Center by standard tests accuracy, constancy, linearity and geometry. To check the performance status of the dose calibrators the activity of radiopharmaceuticals must be accurately known before administered to the patients. Two dose calibrators of Institute Nuclear Medicine and Allied Sciences, Dhaka (Capintec, Inc.,Model no. CRC 25R and Veenstra , model no. VDC 404) performances were compared in accordance with varying geometry because most common measurement geometries utilized in clinical practice are glass vials, solid capsules and various types of plastic syringes. Vials and syringes filling volumes can be widely different, according to clinical needs. It was found that the measurement activity readings of one of these two dose calibrator, VDC-404, showed slightly deviating from accurate value because readings might be affected by the frequent change in background count. The results of the quality control (QC) tests revealed that the parameters that were traced for both dose calibrators were within the limits of the international standards, ?5%. Regular QC should cover precision, accuracy, linearity, and geometry of instrument, according to IAEA standards, e.g. (IAEA TECDOC-602 and 1599), that will guarantee to assure the proper practice of nuclear medicine diagnosis. It is also vital to fulfill radiation protection requirements as well as successful therapies and good quality image production in nuclear medicine practices.

Authors and Affiliations

A. K. M. Mizanur Rahman, S. Reza, Mahfuza Begum, F. Reaz

Keywords

Related Articles

NEST CHARACTERISTICS AND NESTING SUCCESS OF THE INDIAN PIED MYNA (GRACUPICA CONTRA) IN JAMMU REGION (J&K).

The nest characteristics and nesting success of the Indian Pied Myna (Gracupica contra) were studied from May, 2013 to September, 2015 in Jammu region of J&K State. A total of 55 nests were studied. Nesting started in Ap...

PERFORMANCE ANALYSIS OF ACCREDITED COMMUNITY HEALTH CENTRE BASED ON BALANCED SCORECARD WITH STRUCTURAL EQUATION MODELLING IN LUWU TIMUR REGENCY.

The study was conducted at Malili, Mangkutana and Kalaena Community Health Centre in Luwu Timur Regency, South Sulawesi, Indonesia. It aims to analyze the influence of Balanced Scorecard method to measure the performance...

CARICA PAPAYA: A BOTANICAL TREATMENT FOR PATIENTS WITH DENGUE VIRUS INFECTION.

Dengue is a mosquito borne viral disease caused any one of four closely related but antigenically distinct flavovirus serotypes, DENV 1 through 4. Its spectrum ranges from asymptomatic infection to dengue fever (DF), den...

PSYCHOLOGICAL BARRIERS TO LEARNING EFL LISTENING AND SPEAKING AT THE HSC LEVEL: A STUDY IN DHAKA DIVISION, BANGLADESH.

The study aimed at exploring the psychological factors affecting EFL listening and speaking at the HSC level. The researcher primarily reviewed extensive literature in the respective field and finally selected 12 HSC stu...

DESIGN AND PERFORMANCE TEST OF APPROPRIATE ON-FARM COMPOSTING PACKAGE FOR SMALL FARMERS.

In Indonesia, the active participation of farmers is fundamental to gain the multiple benefits of organic waste treatment via composting. Currently, small farmer involvement in composting practice has still low because o...

Download PDF file
  • EP ID EP536116
  • DOI 10.21474/IJAR01/8632
  • Views 59
  • Downloads 0

How To Cite

A. K. M. Mizanur Rahman, S. Reza, Mahfuza Begum, F. Reaz (2019). STUDY OF SOME ESSENTIAL QUALITY CONTROL TEST PARAMETERS FOR THE PERFORMANCE OF DOSE CALIBRATORS IN NUCLEAR MEDICINE PRACTICES.. International Journal of Advanced Research (IJAR), 7(3), 279-291. https://europub.co.uk./articles/-A-536116