Study the Quality Assurance of Conventional X-Ray Machine using Non-Invasive KV Meter

Journal Title: UNKNOWN - Year 2015, Vol 4, Issue 3

Abstract

This study was carried out to obtain optimum operation conditions for X-ray machines. We investigated some factors affecting on quality assurance of conventional Siemen X-ray machine, in one of Mansoura Hospitals such as reproducibility of dose output, time and applied high voltage, Kilo-voltage accuracy, mA accuracy, time accuracy, and linearity. We measured these factors using Non-Invasive kilo voltage meter, The NERO Max 8000 connected with suitable ionization chambers that located at 100 cm source to image detector. Reproducibility of dose output was ranged from 0.1 to 0.7 %, of time was ranged from 0.2 to 3.1 and of high voltage was ranged from 0.1% to 0.7% which is lower than the tolerance limit of the American Association of Physicist in Medicine reference values. Kilo-voltage accuracy percentage was ranged from 1.5 to 3.5 % and time accuracy percentage was ranged from 0.5 to 4.1 % respectively. This study concluded that as the kilo-voltage increases by one, the dose increases by 28%.

Authors and Affiliations

Keywords

Related Articles

Latent and Reactivated Toxoplasma Gondii Infection in HIV-Infected Patients in Albania

"Toxoplasmosis, which is caused by Toxoplasma gondii, is one of the most common parasitic infections of humans and warm blooded animals. The Toxoplasma Encephalitis (TE) is one of the most frequent opportunist infections...

Preparation of Briquettes Using Biomass Combinations and Estimation of Its Calorific Value

"An experimental research design was adopted to conduct the present investigation. For the present study six biomass materials namely Charcoal Dust, Saw dust Rice Husk, Dry Leaves, Wood Chips, Groundnut Shells and two b...

On a New Class of Uniformly Convex Univalent Functions with Negative Coefficient Defined by a Linear Operator

The main purpose of this paper is to introduce new class of uniformly convex functions with negative coefficients defined by a linear operator in theopen unit disc U. We obtain some geometric properties, likes coefficien...

SVM: The Qualitative and Quantitative Monolithic Predictor

Machine Learning is one of the major popular research topics of Artificial Intelligence and its relay with the evolution of techniques and methods which enable the data processor to learn and execute activities. Support...

Crowdedness Spot Acquisition by Using Mobility Based Clustering

Detecting hot spots of moving vehicles in an urban area is absolutely required to many smart city applications. Crowdedness spot is a crowded area with a number of irregular objects. The practical investigation on hot sp...

Download PDF file
  • EP ID EP356927
  • DOI -
  • Views 63
  • Downloads 0

How To Cite

(2015). Study the Quality Assurance of Conventional X-Ray Machine using Non-Invasive KV Meter. UNKNOWN, 4(3), -. https://europub.co.uk./articles/-A-356927