Verification of Radiation Fluence using Stack Film in HDR Brachytherapy with Heterogeneity Algorithm

Journal Title: Journal of Clinical and Diagnostic Research - Year 2018, Vol 12, Issue 11

Abstract

ABSTRACT Introduction: It is essential that Quality Assurance (QA) need to be in line with progresses in HDR brachytherapy treatment planning and delivery to ensure an appropriate level of dosimetric accuracy and quality. Aim: To verify radiation fluence using stack films in HDR brachytherapy with heterogeneity algorithm. Materials and Methods: Gafchromic EBT2 films were stacked in different experimental setup and each film acted as dosimeter to validate AcurosTMBV algorithm in HDR brachytherapy. These films were read using Epson expression 10000 XL flatbed scanner, calibrated against the standard value and analysed with Film QA Pro 2015 software. Different criterion were set for gamma analysis which include Dose Difference (DD)/Distance-To-Agreement (DTA); 2%/1 mm, 5%/1 mm, 10%/1 mm. Results: In order to validate the point doses calculated using model based dose calculation algorithm (AcurosTM BV), a virtual phantom created in the TPS, which agreed with MC based calculation mostly within 2%. The results of each plane were analysed in all the experimental setup. In the gamma pass criteria of 2% and 1 mm, 92% passed, in 5% and 1 mm criteria 96% passed and with 10% and 1 mm criteria, almost 100% passing rate was achieved. Conclusion: Methods provide a comprehensive verification for commissioning the MBDCA-AcurosTM BV and are recommended to use stack film for QA program in HDR brachytherapy.

Authors and Affiliations

Mourougan Sinnatamby, Vivekanandan Nagarajan, Reddy Sathyanarayana Kanipakam, Gunaseelan Karunanidhi, Vijayaprabhu Neelakandan, Saravanan Kandasamy

Keywords

Related Articles

Neoumbilicoplasty in a Laparoscopic Port Site: Description of a New Technique and Review of Literature

The umbilicus contributes significantly to the cosmetic appearance of the abdomen especially in women. Loss of umbilicus may result not only in cosmetic disfigurement but also in significant psychological effects. Omphal...

Correlation of Vascular and Inflammatory Index in Oral Pyogenic Granuloma and Periapical Granuloma – An Insight into Pathogenesis

Introduction: Angiogenesis is vital in the aetiology and pathogenesis of a number of pathological processes that include solid reactive lesions like pyogenic granuloma and chronic inflammatory disorders like periapical g...

Isolated Traumatic Chronic Anterior Dislocation of Radial Head

Isolated traumatic anterior radial head dislocation is extremely rare and very commonly missed injury. A 12-year-old male had a fall on outstretched hand with pronated forearm. He presented after 6 weeks with complaints...

Efficacy of Dexmedetomidine in two Different Doses as an Adjuvant to Lignocaine in Patients Scheduled for Surgeries under Axillary Block

Introduction: The effect of adding dexmedetomidine (α2- adrenoceptor agonist) to local anaesthetics in brachial plexus blocks has been evaluated, however there has been no consensus regarding the ideal dose of dexmedetom...

Overview of the Therapeutic Effects of Origanum vulgare and Hypericum perforatum Based on Iran’s Ethnopharmacological Documents

Herbs have played an important role in the health and wellness of human beings. Nowadays, medicinal herbs are at the forefront of studies of medical science because of their importance in public health. Understanding the...

Download PDF file
  • EP ID EP526283
  • DOI 10.7860/JCDR/2018/36733.12258
  • Views 76
  • Downloads 0

How To Cite

Mourougan Sinnatamby, Vivekanandan Nagarajan, Reddy Sathyanarayana Kanipakam, Gunaseelan Karunanidhi, Vijayaprabhu Neelakandan, Saravanan Kandasamy (2018). Verification of Radiation Fluence using Stack Film in HDR Brachytherapy with Heterogeneity Algorithm. Journal of Clinical and Diagnostic Research, 12(11), 5-10. https://europub.co.uk./articles/-A-526283