What can we Expect from the ORTOJ?

Journal Title: Orthopedics Research and Traumatology – Open Journal - Year 2016, Vol 1, Issue 1

Abstract

The biomechanical performance of a locking compression plate system was compared to an intramedullary photodynamic bone stabilization system in a femur and humerus osteotomy model. The photodynamic bone stabilization system utilizes an angioplasty-like balloon that is introduced into the intramedullary canal of a fractured bone, filled with monomer that is then polymerized and hardened by visible blue light delivered through an optical fiber. This system has been in clinical use since 2010. Synthetic bones engineered to mimic the biomechanical properties of natural bone were cut to produce a 10 mm defect mid-shaft, and two groups of specimens were stabilized by either the compression plate or intramedullary photodynamic bone stabilization system. For each bone model, one locking compression plate system was used, and three different diameter intramedullary photodynamic bone stabilization implants were used. Experimental groups were tested for stiffness, peak load, yield load, peak displacement and yield displacement when a load was applied. Additional samples per experimental group were tested for long-term dynamic stability by cyclically loading until failure. It was found that in all biomechanical parameters measured, the 17 mm intramedullary photodynamic bone stabilization system exceeded the mechanical strength and durability of the locking compression plate system in the femur osteotomy model. It was found that in all biomechanical parameters measured, the 15 mm intramedullary photodynamic bone stabilization system performed equivalently or exceeded the mechanical strength and durability of the locking compression plate system. This testing combined with long-term clinical use, and in vivo data from a large animal model, suggest that femur fixation by an intramedullary photodynamic bone stabilization system will provide equivalent biomechanical properties to a locking compression plate once implanted.

Authors and Affiliations

Samir Shaheen

Keywords

Related Articles

Muscle Pedicle Graft in Avascular Necrosis Femoral Head

Aims: To assess the effectiveness of muscle pedicle graft in avascular necrosis (AVN) head femur in Grade 1 and Grade 2 AVN of femoral head grade 1 and grade 2 as head sparing procedure. Method: The pixel count in Tc-99...

Evaluation of the Ossification of the Cervical Posterior Longitudinal Ligament Utilizing X-Ray, CT and MR Imaging

Background: Cervical ossification of posterior longitudinal ligament (OPLL) is a common disease among the elderly population of East Asian countries. Cervical OPLL is often misdiagnosed in lateral radiography or magnetic...

A Biomechanical Comparison of a Novel Expandable Photodynamic Intramedullary System to a Metal Plate and Screw System in Humerus and Femur Osteotomy Models

The biomechanical performance of a locking compression plate system was compared to an intramedullary photodynamic bone stabilization system in a femur and humerus osteotomy model. The photodynamic bone stabilization sys...

Coaching Orthopaedic Surgeons: Can Visiting Professors be a Valuable Surgical Coach?

Much has been discussed about coaching on different fields and medical training has not been left aside. Atul Gawande’s article in The New Yorker back in 2002 was an eye opener for many regarding the fact that surgeons n...

Os Odontoideum: Review Article

Os odontoideum (OO) is defined as an ossicle with circumferential smooth margins and no osseous continuity with the body of axis. While the size of the ossicle is variable, it has a smooth surface, therefore, the cause o...

Download PDF file
  • EP ID EP552494
  • DOI 10.17140/ORTOJ-1-e002
  • Views 120
  • Downloads 0

How To Cite

Samir Shaheen (2016). What can we Expect from the ORTOJ?. Orthopedics Research and Traumatology – Open Journal, 1(1), 5-6. https://europub.co.uk./articles/-A-552494