Internal Quality control Monitoring & Evaluation of Creatinine, Urea, Alanine aminotransferase and Inter-Laboratory Assessment of Some Haematological Parameters In Two Hospitals in Central Nigeria.

Journal Title: IOSR Journal of Applied Chemistry (IOSR-JAC) - Year 2018, Vol 11, Issue 10

Abstract

The integrity of quality control samples is important to both management of overall quality as well as to meeting requirements of proficiency testing.The study was done to assess the quality control monitoring and evaluation of creatinine,urea,alanine aminotransferaseand also for inter-laboratory studies of somehaematological parameters packed cell volume (PCV),Mean cell Volume (MCV)andmean corpuscular haemoglobin (MCH), in Nisa Premier Hospital Abuja and the Benue State University Teaching Hospital Makurdi,North central Nigeria.The quality control monitoring& evaluation were conducted for two consecutive months using the spectrophotometer(kenza240 TXbiolabo France),and the inter -laboratory studies was performed directly on blood samples at both locations using the Automated Haematology Analyzer (KX-21N Sysmex, USA).The results of quality control assessment showed that Trends (loss of reliability) and shifts (abrupt changes in the control mean) were noted in the results of this study. The results of inter-laboratory assessment of the haematological parameters shows that both imprecision and bias exist between both laboratories. Although imprecision values were generally in the range of “Acceptable to marginal performance”, the overall bias between laboratories was in the range: SDI >1.5 or SDI < -1.5; In this study, Warning signs were recorded but there was no cause for rejection of the analytical run exception of alanine aminotransferase.

Authors and Affiliations

Stephen I. Audu, Isah Jibrin, Maryam Jibrin S. , J. Mohammed, Martha E. Ogbole

Keywords

Related Articles

Synthesis Of Photoconducting Polycarbonate Of Chloroformic ester of 3,6-bis-(hydroxyl methyl)-N-ethyl Carbazole With Bisphenol-A.

N-ethyl carbazole was synthesized by treating Carbazole with ethyl iodide in presence of Sodium hydride in Solvent medium of DMF. N-ethyl-3, 6-diformyl Carbazole was synthesized by reacting N-ethyl carbazole with high ex...

Chemical Analysis of Metal Content In Water of Son River In District Sidhi of madhya Pradesh India

The Present research work undertaken for Chemical Analysis of metal content in water of Son River of district Sidhi during mansoon, winter and summer season in the year 2016-17 where metals such as calcium, magnesium, po...

Nickel (II)-complex ligand as host for Potassium ion guest. Spectroscopic characterization and X-ray structure determination

The title complex, {[Ni(hab)]K[(hab)Ni]}·SCN (H2hab is bis(2-hydroxy-3-methoxybenzylidene)-1,2- diaminobenzene) crystallizes in the Monoclinic space group P21/c with unit cell dimensions a = 11.5891 (13) Å, b = 17.307 (2...

Oxidative cleavage of alpha amino acids ℓ–glycine and ℓ–valine by1,3-dichloro-5,5-dimethylhydantoin in medium of aqueous acetic acid medium

1,3-dichloro-5,5-dimethylhydantoin (DCDMH) hydrolysed of this compounds leads to the formation of active positive halogen species like HOCl and H2O +Cl. The H2O +Cl in non-polar or less polar media, extensively used as a...

Physico-chemical characteristics of River Ganga at Chandighat in Haridwar

In the present investigation physico-chemical characteristics of river Ganga at Chandighat in Haridwar was studied during the period November 2008 to October 2009. Parameters studied were Temperature, pH, free CO2, D.O.,...

Download PDF file
  • EP ID EP414410
  • DOI 10.9790/5736-1110010105.
  • Views 164
  • Downloads 0

How To Cite

Stephen I. Audu, Isah Jibrin, Maryam Jibrin S. , J. Mohammed, Martha E. Ogbole (2018). Internal Quality control Monitoring & Evaluation of Creatinine, Urea, Alanine aminotransferase and Inter-Laboratory Assessment of Some Haematological Parameters In Two Hospitals in Central Nigeria.. IOSR Journal of Applied Chemistry (IOSR-JAC), 11(10), 1-5. https://europub.co.uk./articles/-A-414410