Methodology for Improving Mathematical Model of Ultrasonic Flowmeter to Study Its Error at Distorted Flow Structure

Journal Title: Energy Engineering and Control Systems - Year 2015, Vol 1, Issue 1

Abstract

The paper presents a methodology for improving mathematical model of ultrasonic flowmeter (USM) results through the application of CFD simulation along with experimental reference data on measured consumption. On the basis of the proposed methodology, USM errors research method in terms of distortion of the flow structure has been developed. Using the proposed methodology mathematical model of two-path chordal USM has been improved and its study in terms of distortion of the flow structure after seven types of fittings has been conducted. According to the research results, concrete recommendations on the location of the USM towards the considered fittings have been proposed. The proposed recommendations will improve the accuracy of flow measurement by two-path USM by eliminating additional error due to the presence of distortions in flow structure. Testing of the proposed technique proves its correctness and application for any type of USM and various types of fittings.

Authors and Affiliations

Fedir Matiko, Vitalii Roman, Roman Baitsar

Keywords

Related Articles

Evaluation of Influence of Wall Slip on Measurement of Rheological Parameters by Means of a Hydrodynamic Measuring System

The technique for processing the output signals of hydrodynamic measuring system of rheological parameters of non-Newtonian fluids is presented in the paper. Hydrodynamic measuring system is built on four serially connec...

Techniques for Natural Gas Physical Properties Definition for Flow Rate and Volume Metering Systems

The results of researching the techniques for natural gas properties determination for flow rate and volume metering systems are presented in the paper. The necessity of developing techniques for calculating the isentrop...

Research and Analysis of Rapidly Changing Gas Flow Temperatures Measurement Methods

The temperature of a transducer sensitive element, as a result of its heat capacity, will always lag behind the temperature of the gas flow if this temperature changes. When measuring the flow temperature that varies in...

Measurement of Roughness of Internal Surface of Measuring Pipeline in Real Time for Standard Orifice Plates

Methods for measuring the roughness of the internal surface of a measuring pipeline for projected rectilinear sections of a measuring pipeline and for rectilinear sections that are in operation have been determined. The...

Virtual Source of Reactive Power in Electricity Supply Systems of Household Consumers

It is proposed to include the principle of distributed generation of reactive power in the basic principles of constructing smart microgrids for non-commercial facilities. Individual reactive power compensation devices s...

Download PDF file
  • EP ID EP253648
  • DOI 10.23939/jeecs2015.01.063
  • Views 92
  • Downloads 0

How To Cite

Fedir Matiko, Vitalii Roman, Roman Baitsar (2015). Methodology for Improving Mathematical Model of Ultrasonic Flowmeter to Study Its Error at Distorted Flow Structure. Energy Engineering and Control Systems, 1(1), 63-70. https://europub.co.uk./articles/-A-253648