Phototherapy Radiometer with AS7262 Sensor
Journal Title: Journal of Electronics, Electromedical Engineering, and Medical Informatics - Year 2019, Vol 1, Issue 1
Abstract
The phototherapy is a device used in hyperbilirubinemia therapy by using blue light radiation with ranges between 425nm-475nm. The effectiveness of hyperbilirubinemia therapy depends on the amount of energy emitted by light which expressed in μW/cm2. The purpose of this study is to develop a low-cost and high accuracy Phototherapy radiometer. Measurement of blue light irradiance using the AS7262 sensor which can measure the irradiance of visible light with a wavelength of 450nm, 500nm, 550nm, 570nm, 600nm, 650nm with relative responsiveness of 1 time at each wavelength. SD card memory is used to save measurement data of irradiance so that it can be processed later. Based on the blue light irradiance data collected the smallest error value is 0,40% at a distance of 10cm while the biggest error value is 9,01% at a distance of 30cm. After testing the entire system, the device can be used according to its function and purpose
Authors and Affiliations
Ichwan Syahrul Bahtiar, Andjar Pudji, I Dewa Gede Hari Wisana
The Design of Oxygen Concentration and Oxygen Flow Rate in Continuous Positive Airway Pressure
Measuring oxygen concentration and measuring the flow of oxygen is a tool used to measure the percentage of oxygen content and oxygen flow rate in CPAP. This tool uses the OCS-03F sensor, with Arduino NANO processors the...
Electrosurgery Unit Monopolar Equipped with Cutting and Coagulation Function
Electrosurgery Unit is a medical device that utilizes high frequency and voltage used to cut and dry tissue during the surgical process. The purpose of making this tool is to damage certain body tissues by heating the ti...
Tissue Processor Based Programmable Logic Control
Tissue Processor Tissue Processor consists of several stages of dehydration, clearing, and paraffin infiltration. Phase dehydration to remove water content in tissues by immersion into alcohol. Clearing stage is the proc...
Phototherapy Radiometer with AS7262 Sensor
The phototherapy is a device used in hyperbilirubinemia therapy by using blue light radiation with ranges between 425nm-475nm. The effectiveness of hyperbilirubinemia therapy depends on the amount of energy emitted by li...
Waterbath Design equipped With Temperature Distribution Monitor (Temperature and Timer Control Parameters)
Waterbath is a device used to create a constant temperature. This tool is used to incubates in microbiology analysis. Temperature is maintained according to the desired range. The heating element is controlled by the hea...