RP – HPLC METHOD DEVELOPMENT AND VALIDATION FOR THE ANALYSIS OF METHYL PHENIDETE HYDROCHLORIDE IN PHARMACEUTICAL DOSAGE FORMS

Journal Title: Journal of Atoms and Molecules - Year 2011, Vol 1, Issue 1

Abstract

A simple, selective, linear, precise and accurate RP-HPLC method was developed and validated for rapid assay of Methyl phenidete hydrochloride in tablet dosage form. Isocratic elution at a flow rate of 1ml min-1 was employed on a symmetry C18 column at ambient temperature. The mobile phase consisted of Methanol:acetonitrile:THF 75:20:05 (v/v/v). The UV detection wavelength was at 278nm.Linearity was observed in concentration range of 20-55ppm. The retention time for Methyl phenidete hydrochloride was 3.32 min. The method was validated as per the ICH guidelines. The proposed method can be successfully applied for the estimation of Methyl phenidete hydrochloride in pharmaceutical dosage forms.

Authors and Affiliations

Pammi Kalpana, Syed Reshma, B Sandhya

Keywords

Related Articles

A SENSITIVE DIFFERENTIAL PULSE ADSORPTIVE STRIPPING VOLTAMMETRIC METHOD FOR THE DETERMINATION OF FENOMEDONE

A sensitive differential pulse adsorptive stripping voltammetric method for the determination of fenomedone with universal buffer solution was described. The method was based on the adsorptive accumulation of fenomedone...

RP – HPLC METHOD DEVELOPMENT AND VALIDATION FOR THE ANALYSIS OF TENOFOVUIR IN PHARMACETICAL DOSAGE FORMS

A simple, selective, linear, precise and accurate RP-HPLC method was developed and validated for rapid assay of TENOFOVUIR in tablet dosage form. Isocratic elution at a flow rate of 1ml min-1 was employed on a symmetry C...

ANTIBIOGRAM OF TRADITIONAL MEDICINAL PLANT CALOTROPSIS GIGANTEA EXTRACTS AGAINST STAPHYLOCOCCUS AUREUS AND VIBRIO CHOLERAE.

The increasing incidences of microbial infection and gradual rise in resistance in microbes and the available antibiotics had high-lightened the need to find more alternative antifungal and antimicrobial agents from othe...

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF RUXOLITINIB IN FORMULATIONS

A simple, precise and accurate RP-HPLC method was developed and validated for rapid assay of Ruxolitinib in tablet dosage form. Isocratic elution at a flow rate of 1ml/min was employed on a symmetry Chromosil C18 (250x4....

Download PDF file
  • EP ID EP366871
  • DOI -
  • Views 100
  • Downloads 0

How To Cite

Pammi Kalpana, Syed Reshma, B Sandhya (2011). RP – HPLC METHOD DEVELOPMENT AND VALIDATION FOR THE ANALYSIS OF METHYL PHENIDETE HYDROCHLORIDE IN PHARMACEUTICAL DOSAGE FORMS. Journal of Atoms and Molecules, 1(1), 64-73. https://europub.co.uk./articles/-A-366871