To Evaluate Comparative Values of Schiotz and Applanation Tonometry on Patients of Different Age Groups

Journal Title: IJSR-International Journal Of Scientific Research - Year 2016, Vol 5, Issue 10

Abstract

TITLE: To evaluate comparative values of schiotz and applanation tonometry on patients of different age groups.PURPOSE:To Compare two different methods of intra ocular pressure measurement regarding reliability,accuracy and reproducibility of given methods for screening purpose. MATERIALS AND METHODS:From the out patient attendance of DEPARTMENT OF OPHTHALMOLOGY in PANDIT DEENDAYAL UPADHYAY GOVT. HOSPITAL AND MEDICAL COLLEGE,RAJKOT,300 patients were selected at random for this study.Examination of each patient included slitlamp biomicroscpe and direct ophthalmoscopic examination. Tonometric readings were taken by Goldmann's applanation tonometer, Schiotz tonometer.RESULT AND DISCUSSION:In this study the comparative data obtained on applantion and schiotz tonometer were compared.Schiotz tonometry under estimates the pressure ranging between 16-30 mmHg and overestimates the pressure ranging between 0-15 mmHg.Schiotz tonometer cause decrease in readings of intraocular pressure due to tonographic effect,while multiple readings have no effect on readings taken by goldmann’s applanation tonometer.CONCLUSION:Goldmann’s applanation tonometry is the more reliable,reproducible, method of measuring intra ocular pressure than schiotz tonometer.

Authors and Affiliations

DR. VIMAL J. VYAS, DR BHOLESH RATNA, DR HIRVA P CHAUHAN, DR HARDIK D. JIVANI

Keywords

Related Articles

CLINICAL STUDY ON ECTOPIC PREGNANCY IN GOVERNMENT MEDICAL COLLEGE AND ASSOCIATED GROUP OF HOSPITALS KOTA

Ectopic pregnancy has always challanged ingenuity of the obstetrician and gynecologist by its bizzare clinical picture. Diagnosis is frequently missed which can be life threatning and if not attented in time it can be fa...

Analysis of cases of Endonasal dacryocystorhinostomy(DCR) performed at a tertiary eye care centre.

Purpose of study is to describe technique, success rate of endonasal DCRs and to report complications. Prospective, interventional consecutive study was performed at a tertiary eye care center. Total 40 patients were ope...

Estimation of Height From Footlength : A Scientific Study in the Urban Population of Uttar Pradesh

The present study was done on 150 medical students between the age group of 21 to 30 yrs to find out the correlation of footlength to height and to find whether there were any sexual differences in the urban population o...

AWARENESS OF MEDICAL STUDENTS AND INTERNS ABOUT RECENT CHANGE IN RNTCP GUIDELINE IN TUBERCULOSIS DIAGNOSIS AND MANAGEMENT

Tuberculosis control can signicantly be increased if more effort is given to improve knowledge and attitudes of physicians towards prevention and control of disease. This study aimed to assess the knowledge about recent...

IN VITRO ACTIVITY OF COLISTIN AGAINST MULTIDRUG RESISTANT GRAM NEGATIVE BACILLI (MDR GNB) ISOLATED FROM VARIOUS CLINICAL SPECIMENS

Introduction: Currently colistin is increasingly being used against multidrug resistant gram negative bacteria which include Pseudomonas aeruginosa,,Acinetobacter baumanii, Escherichia coli, Klebsiella pneumoniae & Salmo...

Download PDF file
  • EP ID EP341373
  • DOI -
  • Views 77
  • Downloads 0

How To Cite

DR. VIMAL J. VYAS, DR BHOLESH RATNA, DR HIRVA P CHAUHAN, DR HARDIK D. JIVANI (2016). To Evaluate Comparative Values of Schiotz and Applanation Tonometry on Patients of Different Age Groups. IJSR-International Journal Of Scientific Research, 5(10), 1-3. https://europub.co.uk./articles/-A-341373